a. Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.
b. Place the test cassette on a clean and level surface. Hold the dropper vertically and transfer 1 drop of serum or plasma (40μl) to the specimen well of the test cassette, then add 2 drops of buffer (80μl) * and start the timer. Avoid trapping air bubbles in the specimen well. See the illustration below.
c. Wait for the colored line(s) to appear. Read the result at 15 minutes, do not interpret the result after 20 minutes.
* NOTE: It should be add one more drop of (approx. 40ul) buffer to the sample well of the test cassette if the specimen runs too slowly.