the death of tissue associated with MI, Myoglobin is one of the first markers to rise above normal levels. The level of Myoglobin increases measurably above baseline within 2-4 hours post-infarct, peaking at 9-12 hours and returning to baseline within 24-36 hours. A number of reports suggest the measurement of Myoglobin as a diagnostic aid in confirming the absence of myocardial infarction with negative predictive values of up to 100% reported at certain time periods after onset of symptoms.
The Myoglobin Rapid Test Cassette (Whole Blood/Serum/Plasma) is a simple test that utilizes a combination of anti-Myoglobin antibody coated particles and capture reagents to qualitatively detect Myoglobin in whole blood, serum or plasma. The minimum detection level is 50ng/mL.
¡¾DIRECTIONS FOR USE¡¿
Allow the test, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it within one hour.
2. Place the cassette on a clean and level surface.
For Serum or Plasma specimen:
· Hold the dropper vertically and transfer 2 drops of serum or plasma (approximately 50µL) to the specimen area, then add 1 drop of buffer (approximately 40µL), and start the timer. See illustration below.
For Venipuncture Whole Blood specimen:
· Hold the dropper vertically and transfer 3 drops of whole blood (approximately 75µL) to the specimen area, then add 1 drop of buffer (approximately 40µL), and start the timer. See illustration below.
For Fingerstick Whole Blood specimen:
· To use a capillary tube: Fill the capillary tube and transfer approximately 75µL of fingerstick whole blood specimen to the specimen area of test cassette, then add 1 drop of buffer (approximately 40µL) and start the timer. See illustration below.
· To use hanging drops: Allow 3 hanging drops of fingerstick whole blood specimen (approximately 75µL) to fall into the specimen area of test cassette, then add 1 drop of buffer (approximately 40 μL) and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. Read results at 10 minutes. Do not interpret the result after 20 minutes.
Cat. No. |
Product Description |
Specimen |
Format |
Kit Size |
Cut-Off |
Status |
CMY-402 |
Myoglobin Rapid Test Cassette |
WB/S/P |
Cassette |
10 T |
50 ng/mL |
CE |